Cut through Test Apparatus

  • Reference Standard: IS 13778 Part-6:2011, IEC 60851-6:1996 and other equivalent standards.
  • Test: Cut through test expressed as the temperature at which a short circuit occurs between two pieces of wire crossing each other at right angles with a specified load applied to the crossing point.
  • Test Voltage: 100 ±
  • Digital AC Voltmeter range 199.9VAC; LC: 0.1V and accuracy Class 0.5
  • When the output terminals are shorted the Current is limited to a maximum of 50mA.
  • Fault Indicator ON when a current of 5 ±
  • Digital AC milli Ammeter range 199.9mA; LC: 0.1mA and accuracy class 0.5
  • Temperature Controller cum Indicator: Range: 400°C, LC: 0.1°C and Accuracy ± (0.5%+1.0°C).
  • Digital Timer (Test Time) Range: 999s; LC: 1s.
  • Force: 4.5 to 36N (4.5-2No’s, 9.0-1No & 18.0N-1No.)
  • Input Supply requirements: Single Phase 230VAC, 50Hz, 6A with proper Earthing.