Reference Standard: IS: 10810 Part-58, IS: 10810 Part-59, ASTM D 2863 and other Equivalent Standards.
Method: Determine the minimum concentration of Oxygen in a mixture of Oxygen and Nitrogen that will just support the combustion of material under equilibrium conditions of “Candle like” burning at Room Temperature FTA and at High-Temperature HFTA.
Direct Readout of %Oxygen on an Analyzer.
Analyzer: Electrochemical OR Magnetic OR Paramagnetic As per Customer requirements/Budget.
Gas Inlet: ON/OFF Valve provided.
Gas Flow: Acrylic Rotameter provided. Individual Rotameter for Oxygen, Nitrogen and also for the total flow.
Gas Control: Micrometer Valve for controlling gas to get fine control.
Temperature Controller: PID Temperature Controller with RAMP setting to avoid temperature shoot-up with PT-100 3 wire Air Sensor.
Test Colum: Double wall with Heat resistance transparent Glass (Toughened Glass). 75mm ID and 450mm long.
Specimen holder with Screen will be provided.
Glass beads will be provided to fill a minimum of 100mm of the test column.
Test applicable for a wide range of materials such as Polymers, Cables, Textiles, Plastics, Rubber, Nylon, etc..
Available with Digital Flow Meters and Digital Flow Controllers also.